    Comes from university surplus, see specs and description from manufacturer below.  Manual is available for pdf download from their website.
    Boxcar Averager
    from 90
    SR250 — Gated integrator with gates to 2 ns
    Gate width from 2 ns to 15 µs
    (expandable to 150 µsec)
    Internal rate generator
    Active baseline subtraction
    Shot-by-shot output
    Gate output for precise gate timing
    Average 1 to 10,000 samples
    DC to 20 kHz repetition rate
    Low jitter
    Overview
    The SR250 Gated Integrator is a versatile, high-speed, NIM module designed to recover fast analog signals from noisy backgrounds. The SR250 consists of a gate generator, a fast gated integrator, and exponential averaging circuitry. The gate generator, triggered internally or externally, provides an adjustable delay from a few nanoseconds to 100 ms before it generates a continuously adjustable gate with a width between 2 ns and 15 µs.
    SR250 Gated Integrator
    The SR250 Gated Integrator is a versatile, high-speed, NIM module designed to recover fast analog signals from noisy backgrounds. The SR250 consists of a gate generator, a fast gated integrator, and exponential averaging circuitry. The gate generator, triggered internally or externally, provides an adjustable delay from a few nanoseconds to 100 ms before it generates a continuously adjustable gate with a width between 2 ns and 15 µs. The gate delay can be set from the front panel or automatically scanned by applying a rear-panel control voltage. Scanning the gate allows the recovery of entire waveforms.
    The fast gated integrator integrates the input signal during the gate. The output from the integrator is then normalized by the gate width to provide a voltage proportional to the average of the input signal during the sampling gate. This signal is further amplified and sampled by a low-droop sample-and-hold amplifier, and output via a front-panel BNC connector. The last sample output provides a shot-by-shot analysis of the signal, and makes the instrument a particularly useful component in a computer data acquisition system.
    Triggering
    The SR250 may be triggered internally or externally. The internal rate generator is continuously variable from 0.5 Hz to 20 kHz in nine ranges. The external trigger pulse may be as short as 5 ns, allowing the unit to be triggered with fast pulses from photodiodes and photomultipliers. Single shot and line triggering can also be selected.
    Signal Inputs
    The sensitivity of the instrument (volts out /volts in) may be set from 1 V/1 V to 1 V/5 mV. If additional gain is required, the SR250 can be used with the SR240 preamplifier. The input is protected to 100 V and has a 1 MΩ input impedance. An input filter rejects unwanted signals before the input is sampled by the integrator. Unwanted DC input offsets are easily nulled with a 10-turn potentiometer.
    Gate Timing
    The delay of the sample gate from the trigger is set by the delay multiplier and scale. The delay scale is multiplied by the setting on the 10-turn multiplier dial, allowing continuously adjustable delays from a few nanoseconds to 100 milliseconds. The delay multiplier may also be changed from the rear-panel control voltage input—a useful feature in applications requiring a scanning gate. Zero to ten volts at this input overrides the front-panel 0 to 10× delay multiplier. Insertion delay from trigger to gate is only 25 ns, and gate-delay jitter is only 20 ps + 0.01 % of the full-scale delay.
    The width of the sampling gate may be continuously adjusted from 2 ns to 15 µs over eight width ranges. A simple modification of the unit allows gate widths of up to 150 µs. The front-panel gate output provides a representation of the gate that can be overlayed with the signal on an oscilloscope to provide a precise display of the gate timing.
    Signal Outputs
    A moving exponential average of 1 to 10,000 samples can be selected from the front panel. This traditional averaging technique is useful for pulling small signals from noisy backgrounds. In the case of a random white noise background, the signal-to-noise ratio increases by the square root of the number of samples in the average. This allows a S/N improvement of up to a factor of 100 using this technique alone. If no averaging is desired, or if averaging is to be performed a computer, the last sample output provides a voltage proportional to the average value of the input signal during the last gate period.
    Average Reset
    The reset button sets the average output to zero. The average may also be reset by a rear-panel logic input. The average reset input will accept a TTL signal or a switch closure to ground to reset the moving average output.
    Polarity Control and Active Baseline Subtraction
    The polarity of the last sample and averaged outputs is controlled by rear-panel toggle switches. Positive outputs can be selected for negative signals, and vice versa, allowing easy interfacing with unipolar analog-to-digital conversion systems. In addition to the traditional averaging modes, the SR250 possesses a unique Active Baseline Subtraction mode which allows you to actively cancel baseline drift. In the Active Baseline Subtraction mode the SR250 is triggered at twice the source repetition rate. On alternate triggers, when the signal is not present, only the baseline is sampled, and the SR250 inverts the polarity of the last sample output before it is added to the moving average. Thus, any baseline drift not associated with the source will be subtracted out.
    Additional Outputs
    The signal input is passed on to the signal output by a length of coaxial cable for termination and for gate timing. It is delayed exactly 3.5 ns from the input, and can be terminated to optimize either signal gain or response time. The gate output provides a pulse synchronized with the internal gate signal. The gate output is timed so that it can be overlayed with the signal output for precise adjustment of gate timing. The busy output provides a TTL timing pulse which is high while the unit is integrating, and goes low when the SR250 is ready to accept another trigger. These outputs help simplify experimental setup and troubleshooting.
    SR250 Gated Integrator Specifications
    Trigger
    Internal trigger
    0.5 Hz to 20 kHz
    Line trigger
    The gate generator may be triggered from AC line with adjustable phase.
    External trigger
    1 MΩ input impedance. Trigger threshold adjustable from 0.5 to 2 V. Input protected to ±100 VDC. Trigger pulse must be over threshold for >5 ns with a rise time <1 µs.
    Manual trigger
    The unit will trigger if trigger threshold is scanned through 0 VDC.
    Trigger LED
    LED blinks with each trigger.
    Delay
    Delay scale
    1 ns to 10 ms
    Delay multiplier
    0 to 10× using 10-turn dial
    Insertion delay
    25 ns
    Accuracy
    2 ns or 5 % of full-scale delay, whichever is larger
    Jitter
    <20 ps or 0.01 % of full-scale delay, whichever is larger
    Ext. delay control
    Rear-panel 0 to 10 VDC input overrides front-panel delay multiplier. Input is used by SR200 or SR245 to scan the gate.
    Gate Width
    Width scale
    1, 3, 10, 30, 100, 300 ns, 1, 3 µs
    Width multiplier
    Adjustable from 1× to 5×
    Width accuracy
    2 ns or 20 % of full scale, whichever is greater
    Minimum width
    2 ns, FWHM
    Signal
    Sensitivity (V
    in
    /V
    out
    )
    1V/V to 5mV/V in a 1-2-5 sequence
    Accuracy
    3 % for gate widths >10 ns, decreasing to 50 % for a 2 ns gate
    Filter
    DC coupled, or AC coupled above 10 Hz or 10 kHz
    Offset control
    ±0.4 VDC using 10-turn dial.
    Over range LED
    Indicates input is >2 VDC or LAST SAMPLE is greater than 10 VDC
    Last Sample
    Output
    ±10 VDC, 10 mA (20 mA short circuit limit), impedance <1 Ω
    Polarity switch
    Inverts LAST SAMPLE output
    Responsivity
    95 % (no more than 5 % of the previous last sample remains)
    Averaging
    Type
    Exponential moving average
    Number of samples
    1, 3, 10, 30, ... to 10,000. LAST is selected for no averaging.
    Average output
    ±10 VDC full scale, 10 mA (20 mA short circuit limit). Impedance <1 Ω
    Droop rate
    When no ext. triggers are present, droop rate is <1 % per minute (1 to 30 samples), and <0.01 % per minute (100 to 10,000 samples).
    Average polarity and baseline subtraction
    Rear-panel switch sets polarity of LAST SAMPLE before it is added to the average. Can also be used to invert polarity of average output. In TOGGLE position, every other sample is subtracted from the average. By triggering at twice the experiment's rep rate, baseline will be sampled on alternate triggers and subtracted from the average.
    Toggle output
    Rear-panel TTL signal changes state with each trigger. Output used with Active Baseline Subtraction feature to indicate if next sample will be added to, or subtracted from, the moving average. Toggle output can drive 50 Ω loads to +2 VDC.
    Reset button
    Resets average to zero
    Remote reset
    Rear-panel input resets average with a TTL low or switch closure.
    Signal Input and Output
    Signal input
    1 MΩ input impedance, ±2 VDC usable range, protected to 100 VDC. Input offset drift <0.5 mV/hr. after 20 minute warm-up. Shot noise at input <0.5 mV. Coherent pickup <5 mV (easily canceled with offset knob in fixed gate applications).
    Signal output
    SIGNAL OUTPUT is the input signal delayed by 3.5 ns. (Used to terminate input signal and to time gate with respect to signal output.)
    Gate and Busy Outputs
    Gate output
    200 mV pulse marks exact position of gate with respect to signal output. ±1 ns accuracy (50 Ω load)
    Busy output
    TTL signal indicates output data is ready. Stays high from trigger until unit is ready for next trigger. (45 µs min., longer for long delays or gate widths.) Drives 50 Ω load to 2 VDC.
    General
    Power supplies
    +24V/135 mA, +12V/380 mA, -12V/230 mA, -24V/150 mA. 14 W. Power from a standard NIM crate (SR280).
    Mechanical
    Dual-width NIM enclosure
    Dimensions
    2.7" × 8.174" × 11.5" (WHL)